Chapter 5 Error Messages
Calibration Errors
122
713 Bad readback cal data
The specified readback calibration values (
CAL:VOLT
or
CAL:CURR
) are out
of range. Note that the new calibration constants are not stored in the non-
volatile memory.
714 Bad OVP cal data
The overvoltage protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
717 Cal OVP status enabled
Overvoltage protection status is enabled. You must set overvoltage to OFF
before and during the calibration.
718 Gain out of range for Gain Error Correction
The slope of the DAC gain is out of range. Hardware fails.
740 Cal checksum failed, secure state
741 Cal checksum failed, string data
743 Cal checksum failed, store/recall data in location 1
744 Cal checksum failed, store/recall data in location 2
745 Cal checksum failed, store/recall data in location 3
746 Cal checksum failed, DAC cal constants
747 Cal checksum failed, readback cal constants
748 Cal checksum failed, GPIB address
749 Cal checksum failed, internal data
754 Cal checksum failed, store/recall data in location 4
755 Cal checksum failed, store/recall data in location 5