172 FONIX FP35 Hearing Aid Analyzer
13. Curve ID box. See Section 2.7.2 for more information.
14. Source Signal Statistics Box. The information in the box is dependent upon
signal type.
• COMP,DIGSPCH—DisplaystheRMSsourceamplitude.
• NORM,FAST,SHORT,LONG,SINGLE—Displaysthesourceamplitude,
frequency, distortion, and output/gain.
Fig 6.2.2—Coupler EarSim screen displayed with simulated insertion gain
6.3 Measurements & Adjustments
Taking measurements in the Coupler Target and the Coupler EarSim screens is
just like taking measurements in the Coupler Multicurve screen. The only differ-
ence is the addition of the target (and possibly audiometric data) on the screen.
Any measurements taken in one of these screens will show up in the other
screens. For a detailed explanation of the measurements available in all coupler
measurement screens, see Chapter 3.
The Coupler Multicurve and Coupler Target screens will show identical mea-
surements. The Coupler EarSim screen converts these measurements into simu-
lated real-ear measurements, so their numerical values will be different than
those displayed in the other coupler screens.