Specifications
Specifications
The "supplemental information" and "typical" entries, in the following
specifications are not warranted, but provide useful information about the functions
and performance of the instruments.
The following specifications data is specified at 23 ± 5 °C and 50 % relative
humidity.
Function
Agilent 16440A switches either a SMU or PGU to the associated output port.
You can expand to 4 channels by adding an additional 16440A. The channel 1
PGU port provides "PGU OPEN" function, which can disconnect the PGU by
opening a semiconductor relay. The 16440A can not work without two pulse
generator units of Agilent 41501 SMU and Pulse Generator Expander.
Channel configuration: 2 channels (CH1 and CH2). Can add additional 2
channels (CH3 and CH4) by adding another 16440A (selector expander).
Voltage and current range
Input Output
Channel 1 (CH 1) 2 (SMU and PGU) 1
Channel 2 (CH 2) 2 (SMU and PGU) 1
Channel 3 (CH 3)
a
a. These channels are available when an 16440A SMU/PG selector
expander is installed.
2 (SMU and PGU) 1
Channel 4 (CH 4)
a
2 (SMU and PGU) 1
Input port Maximum Voltage Maximum Current
SMU 200 V 1.0 A
PGU 40 V
0.2 A
a
a. This is peak-to-peak ac current.