Agilent Technologies 667xA Welding System User Manual


 
38 Language Dictionary
Initiate Subsystem
INIT
INIT:CONT
This subsystem enables the trigger system. When a trigger is enabled, an event on a selected trigger source causes the
specified trigging action to occur. If the trigger subsystem is not enabled, all trigger commands are ignored. If
INIT:CONT is OFF, then INIT enables the trigger subsystem only for a single trigger action. The subsystem must be
enabled prior to each subsequent trigger action. If INIT:CONT is ON, then the trigger subsystem is continuously enabled
and INIT is redundant.
Command Syntax INITiate[:IMMediate]
INITiate:CONTinuous <bool>
Parameters For INIT[:IMM] (None)
For INIT:CONT 0|1|OFF|ON
*RST Value OFF
Examples INIT INITIATE:IMMEDIATE
INIT:CONT 1 INITIATE:CONTINUOUS 1
Query Syntax For INIT[:IMM] (None)
For INIT:CONT INIT:CONT?
Returned Parameters <NR1> 0|1
Related Commands ABOR <GET> *RST TRIG *TRG
Measure Subsystem
MEAS:CURR?
MEAS:VOLT?
This query subsystem returns the voltage and current measured at the power supply's sense terminals.
Query Syntax MEASure:CURRent[:DC]? MEASure:VOLTage[:DC]?
Parameters (None)
Default Suffix A for MEAS:CURR? V for MEAS:VOLT?
Examples MEAS:CURR? MEAS:VOLT? MEASURE:VOLTAGE:DC? MV
Returned Parameters <NR3>