Page 6
Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator
AC Voltage Range: 0-750V AC
Accuracy: ±0.07% (Basic)
Frequency Range: 40 ~ 500Hz
Accuracy: ± 2%
DC Voltage Range: 0-1000V DC
Accuracy: ±0.006% (Basic)
Current Range: 0 ~ 100A
Accuracy: AC (40Hz to 100Hz): ±0.08% (Basic)
Resistance Range: 1Ω ~ 2000Ω
Accuracy: ±0.03% (Basic)
Fluke 5500 or Wavetek
9100 Calibrator or
Equivalent
Ballantine 1620A
Form number TM61095 Rev 2 September 2002