Invacare Compressor Air Compressor User Manual


 
SECTION 5—INDICATOR LIGHTS
Part No 1100873 32 HomeFill
II Compressor
SECTION 5—INDICATOR LIGHTS
INDICATOR LIGHT
COLOR
COMPRESSOR
STATUS
EFFECT
None Compressor is Off O
2
flowing to user if
concentrator is on
O
2
BELOW NORMAL -
YELLOW
Compressor is On and
warming up
(0-3 minutes after
power is On)
Cylinder filling not
started
O
2
flowing to user if
concentrator is On
O
2
from concentrator
is < 90% (any time after
three minute warmup)
Cylinder filling stops,
will restart when O
2
> 90%
O
2
flowing to user if
concentrator is On
FILLING - GREEN Compressor is On and
cylinder is filling. O
2
from concentrator is
>90% (any time after
three minute warmup)
Cylinder is filling
O
2
flowing to user if
concentrator is On
FULL - GREEN Compressor is On.
Cylinder is finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not
attached properly -
check connection
(FULL light is On in
<10 mins.)
Cylinder filling stops
Cannot fill another
cylinder until
compressor is turned
Off and On again
O
2
flowing to user if
concentrator is On