SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES
HomeFill
II Compressor 35 Part No 1100873
O
2
BELOW NORMAL
(YELLOW) light stays On
1. Ensure the concentrator has warmed up for at
least 20 minutes.If it has not been 20 minutes,
turn the compressor Off until the concentrator
warms up. If the YELLOW light still remains on
proceed to STEP 2.
2. Inspect the connection between the compres-
sor and the concentrator for damage or pinching.
If damaged, replace. If pinched, straighten tubing.
Turn the compressor Off for at least 10
seconds.
Turn the compressor On. If the YELLOW light
still remains on after 10 minutes, proceed to
STEP 3.
3. The concentrator may need service, contact
your home health care provider or Invacare.
Fill times exceed
maximum as
recommended fill times
OR GREEN
filling light stays On.
1. Ensure the cylinder regulator is set to Off.
2. Check the recommended fill times shown in
Cylinder Fill Times
on page 36.
3. Check that the cylinder is connected properly.
Turn the compressor
Off. Remove the cylinder and reinstall. Turn the
compressor ON.
4. Check the compressor/concentrator line for
proper connection and/or damage (i.e. - leaks or
kinks).
5. Check the concentrator liter flow setting is at
three L/min. or less. If necessary, adjust the liter
flow setting (refer to the concentrator owner’s
manual). If the GREEN filling light does not go Off,
the compressor and/or concentrator may need
service. Contact your home health care
provider or Invacare.
Excessive noise when
compressor is On
1. Call home health care provider or Invacare for
service.
Hissing sound from Regu-
lator when flow selector
is Off
1. Turn flow selector Off.
2. If hissing sounds continue, call home health
care provider or Invacare.
TROUBLESHOOTING
PROBLEM SOLUTION