SECTION 6—TYPICAL PRODUCT PARAMETERS
Invacare® Element™ 20 Part No 1160840
OXYGEN CYLINDER DURATION
Thetotaldeliv eryof oxygenviatheconserverisrelatedtothe
breathingrates,itisuseradaptiv einthatthetotaloxygen
deliveredperminutewillautomaticallyadjustwithuserneed,as
expressedbyincreasedordecreasedbre athingrates.For
example,atallse ttings,twiceasmuch oxygenperminutewillbe
deliveredifonebre athestwenty(20)timesperminuteas
comparedwithte n(10)timesperminute.RefertoOXYGEN
DURATIONTABLEbelow.
OXYGEN DURATION TABLE
SHOCK VIBRATION: Not to exceed IEC 601-1 requirements
Not to exceed IEC 68-2-6, IEC 68-2-34
No smoking or open flames
Consult accompanying documents
SETTING 123456Continuous
Flow
2 L/min.
Cylinder Type Cylinder Volume
Liters @ 2015 psi
Estimated Cylinder Duration in Hours
(Based on 20 breaths/min.)
M2 36 1.9.9.6.5.4.3 .3
M4(A) 113 5.9 2.9 2.0 1.5 1.2 1.0 .9
M6(B) 164 8.5 4.3 2.8 2.1 1.7 1.4 1.4
ML6 171 8.9 4.5 3.0 2.2 1.8 1.5 1.4
M9(C) 246 12.8 6.4 4.3 3.2 2.6 2.1 2.1
D 425 22.1 11.1 7.4 5.5 4.4 3.7 3.5
E 680 35.4 17.7 11.8 8.9 7.1 5.9 5.7
ƽ