SECTION 8—TROUBLESHOOTING
Invacare® Element™ 22 Part No 1160840
SECTION 8—TROUBLESHOOTING
Ifoxygenconservingregulatorfailstofunction,consultthe
Troubleshootingguide.Ifproblemcannotbecorrected,
consultyourDealer/HomeCareequipmentprovider.
Replacement Parts - Accessories
HF2POST6BAGCarryingBagorHF2POST9BAGCarryingBag
PROBLEM POSSIBLE CAUSE SOLUTION
No flow.
Unit does not pulse.
1. Cylinder valve is closed.
2. Regulator in “Off”
position.
3. Cylinder empty.
4. Oxygen conserving
regulator not sensing
breath.
5. Oxygen cannula is
blocked or kinked.
1. Turn the cylinder valve to
the “ON” position.
2. Set to prescribed setting.
3. Replace cylinder.
4. Check position of
cannula in nose. DO
NOT breathe through
mouth.
5. Remove kinks. Clean or
replace, if necessary.
Unit flows or pulses
continuously.
1. Unit is set to the “CF”
position.
2. Unit was not set to
“OFF” prior to opening
the cylinder valve.
3. Vent hole is obstructed.
1. Turn the selector switch
to the appropriate
delivery setting.
2. Turn the selector switch
to “OFF,” wait a few
moments, then set at
proper delivery setting.
3. Remove obstructions,
such as labels or carrying
bag, and resume use as
usual.