Tektronix TDS 500D Sander User Manual


 
Performance Tests
1–36
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Equipment
required
One sine wave generator (Item 14)
One level meter and power sensor (Item 15)
One power splitter (Item 16)
One female N to male BNC adapter (Item 18)
Four male N to female BNC adapters (Item 17)
Two 50 W precision cables (Item 5)
Attenuators (Items 1 and 2)
Optional: One high-frequency leveled sine wave generator and its
leveling head – replaces items 14, 15, 16, 17, and 18
Prerequisites See page 1–15
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
H Press save/recall SETUP. Press the main-menu button Recall
Factory Setup; then press the side-menu button OK Confirm
Factory Init.
b. Modify the default settings:
H Press TRIGGER MENU.
H Press the main-menu button Coupling. Then press the side menu
button Noise Rej.
H Turn the horizontal SCALE knob to 50 ns. Press SHIFT; then press
ACQUIRE MENU.
H Press the main-menu button Mode; then press the side-menu button
Average 16.
H Press MEASURE. Press the main-menu button Level Setup; then
press the side-menu button Min-Max.
Check Analog Bandwidth