Calibrating the 5700A/5720A Series II to External Standards
3
3-11
8. Reverse the HI and LO connections at the 732B terminals, and press E. The
following displays appear, asking you to wait before proceeding with the 10 k
standard.
PREV
MENU
IF3-8.EPS
After a few seconds, the following display appears:
PREV
MENU
IF3-9.EPS
9. Connect the calibrator to the 10 k standard as shown in Figure 3-4 and enter the
true value of the standard. If the standard is not between 9 k and 11 k, an error
message appears, which allows you to start over from this point with a different
standard. Press E again to open the following display:
PREV
MENU
IF3-10.EPS
When the internal 10 k reference has been characterized, the following message
appears, which lets you accept or reject the changes that are about to be made to the
calibration constant:
PREV
MENU
IF3-11.EPS