14
5 ThedisplayreadsPLEASEREPLACEFILTER
Openthereardooroftheunitandunscrewtheold
lterSeeFigure8
6
Lookatthenewlter—verifybotho-ringsare
lubricatedandcorrectlylocatedinthegroovesSee
Figure9ThreadthenewlterintoplacePress
START / YES
Note: If the display shows SERIALNUMBERERROR,
the serial number is invalid or the lter-drier has
already been used in this unit.
7 ClosethedoorPressSTOP / NOtoexitThelter-
drierreplacementisnowcomplete
Replace the Filter-Drier contd.
Maintenance
Filter-Drier
Rear View
(without shroud)
Adjust Tank Fill Level
Whenconnectedtoarefrigerantsource,theunit
maintainsadefaultvalueof15lbsofrefrigerantinthe
internalstoragevesselThisvaluemaybeadjustedup
ordowntosuittheapplication
1PressMENU
2 Usethearrowkeystoscrollthroughthemenuto
ADJUSTTANKFILLLVL,andpressSTART /
YESTheunitdisplays
LEVEL:1500LBS
ENTERTANKFILLLVL
LIMIT:4TO17LBS
PRESSSTART/YESTOSAVE
3 Usethearrowkeystoenteravaluebetween4lbs
and17lbs,andpressSTART / YES.
Figure 8
O-rings
Figure 9
Check for Leaks
Checktheunitforleakseverythreemonths,oras
speciedbylawinyourjurisdiction
1 TurnoffMAIN POWER,anddisconnectthepower
cordfromtheoutlet
2
Openthereardoor;removetopcoverandfrontpanel
3
Usealeakdetectortoprobeallconnectionsfor
refrigerantleaksTightenttingsifaleakisindicated
4 Reassemblethebodypanels,andclosethereardoor
Inspect the unit periodically for leaks.
The manufacturer does not reimburse
for lost refrigerant.