A SERVICE OF

logo

16
Repl.
Component Part No.
Filter-Drier 34724
ServiceCouplerSet
(high- and low-side couplers) 18192
ServiceHoseSet
(high- and low-side hoses) 34722
VacuumPumpOil
(case of 12 quarts) 13203
VacuumPumpOil
(case of 4 gallons) 13204
MaintenanceKit 13172
VinylDustCover(optional) 17492
TankAdapter 16301
Replacement Parts
Maintenance
Scale Calibration
Thecalibrationcheckisusedtoensuretheunit’sinternal
scaleisalwayscalibratedDuringthistest,useonlythe
calibrationweightthatisprovidedwiththeunit
1 PressMenu
2 UsethearrowkeystoscrolltoCALIBRATION
CHECK
3 RefertoFigure11,andverifythemagnetonthe
bottomoftheunitiscleanPressSTART / YES
4 Thedisplayreads
ATTACHTHEWEIGHTTOTHEBOTTOMOFTHE
MACHINE
PRESSSTARTTOCONTINUE
5Attachtheweightprovidedtothemagnetonthe
bottomoftheunitPressSTART / YES
Magnet
• IfthedisplayshowsCALIBRATIONAPPROVED,thescaleisincalibrationPressanykeytoexit
• IfthedisplayshowsCALIBRATIONREJECT,thescaleisoutofcalibrationCallyourlocalservice
centerforassistance
• IfthedisplayshowsCALIBRATIONERROR1,verifythecalibrationweightiscorrectlyattached
tothemagnet,andverifythemagnetiscleanwithnothingattachedtoitIfthatdoesnotcorrectthe
problem,callyourlocalservicecenterforassistance
Tank Recertication
Theinternalstoragevessel(ISV)inthisunitisa
DOT-approvedtankthatcarriesadatecodeISVs
mustberecertiedeveryveyears
TodeterminethedatecodeoftheISVinyour
Robinairunit,removethefrontpaneloftheunitIt
shouldbepossibletoreadthedatecodestampedon
theISV
Ifthedatecodereadsbeyondtheve-yearlimit,
contactaRobinairservicecenterforassistance
Figure 11