85036B/E 1-7
General Information
Calibration Kits Documented in this Manual
Calibration Kits Documented in this Manual
This manual applies to any 85036B or 85036E calibration kit whose serial prefix is listed
on the title page. If your calibration kit has a different serial number prefix than the one
listed on the title page, refer to “Calibration Kit History” below for information on how this
manual applies.
Calibration Kit History
This section describes calibration kits with serial number prefixes different than those
listed on the title page.
Agilent 85036B kits with serial prefix 3102A and below
75 ohm load specification: return loss 45 db, dc to 2 GHz
all other kit components unspecified
85036B calibration kits with serial number prefix 3102A and below should meet the
specifications published in this manual. If you wish to have your kit recertified to current
specifications, contact your nearest Agilent service center. For contact information, see
Table 5-1 on page 5-3.
Agilent 85036B kits with serial prefix 3103A and below
These calibration kits had a center conductor extender with a smaller body. Therefore, a
different procedure was used for connecting the female open. If you’re using one of the
smaller extenders, replace steps 1 through 3 in “Connecting and Disconnecting the
Two-Piece Female Open (85036B)” on page 3-9 of this manual with the steps below.
1. Mate the center conductor extender to the male pin of the test port.
2. Connect the open body to the test port by carefully guiding it over the extender.
3. To disconnect the open, carefully remove the open body first.
The part numbers in this manual are the recommended replacement parts for these kits
(including the center conductor extender).