Agilent Technologies E4445A Saw User Manual


 
Chapter 5 469
One-Button Measurement Functions
Meas Setup (Spectrum Emissions Mask—SEM)
One-Button Measurement Functions
SEM:OFFS:LIST:BAND:IMUL? or SEM:OFFS:LIST:BWID:IMUL?
SEM:OFFS:LIST:BAND:AUTO 0 or SEM:OFFS:LIST:BWID:AUTO 0
SEM:OFFS:LIST:BAND:AUTO? or SEM:OFFS:LIST:BWID:AUTO?
5.22.4.7 Abs Start
Enables you to enter an absolute level limit at
Start Freq for the currently selected offset
ranging from 200.00 to +50.00 dBm with 0.01 dB resolution.
Key Path:
Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for
“Offset/Limits” on page 462 for this value.
Default Terminator: dBm
Range: –200 dBm to 50 dBm
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute <ampl>,...[<ampl>] (up to five
values)
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2
for MS. Default is BTS (1).
Example: SEM:OFFS:LIST:STAR:ABS 20 dBm
SEM:OFFS:LIST:STAR:ABS
5.22.4.8 Abs Stop
Enables you to enter an absolute level limit at
Stop Freq for the currently selected offset
ranging from 200.00 to +50.00 dBm with 0.01 dB resolution, and to toggle this function
between
Couple and Man.IfsettoCouple,thisiscoupledtoAbs Start to make a flat limit line.
If set to Man, Abs Start and Abs Stop you can enter different values to make a sloped limit
line.
Key Path:
Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for
“Offset/Limits” on page 462 for this value.
Default Terminator: dBm