Hach 3650 ATEX Sander User Manual


 
Page 4
Viewing Stored Measurements
Switch the instrument OFF (by pressing the
POWER key). Hold down the × button while
switching the instrument back ON. The LCD
displays a sample location number.
Scroll through the numbered sample locations of
all the stored values using the × Ø buttons.
To view the actual gas concentration
measurement value at a particular sample
number, press the STO button. The LCD now
displays the stored value for that sample number.
Press STO a second time to return to the next
numbered location display, to continue scrolling
or view another stored value.
To return to the measurement mode, switch the
instrument OFF and then back ON again without
holding down any additional buttons.
Downloading Data
If you have made measurements and stored them
in the instrument, you should be ready to bring
them into the WinLog97 program for viewing,
copying, saving and printing. To download stored
data from the instrument to the PC, choose the
DownLoad data command from the Logger
menu.
The window displays the stored measurements
from the instrument showing the sample number,
gas concentration, a date and time stamp, and a
sample description.
Sample Description
For help in identifying the locations of various
sampling points that are stored by the analyzer,
you may choose the Sampling Point
Description command from the Logger menu.
The measurement values to be placed in
positions 0 through 499 can be described
however you wish. Double-click on a particular
position (or click Modify), then type a description
in the box.
Copy Data
To copy the results to the Windows Clipboard, so
that the data can be pasted into a spreadsheet,
word processor or other Windows program that
accepts tabular text information, choose the
Clipboard command from the Export menu.
Save Data
To save this list of measurements as a text file,
capable of being recalled by the WinLog97
program or imported as a file into other Windows
programs, choose the Save As command from
the File menu and enter the device and file name.
Print Data
To place this list of measurements into a tabular
format and send it to the printer, choose the Print
command from the File menu.
Clear Data
To clear all the values stored in the instrument,
choose the Clear Data command from the
Logger menu. Since this action will clear the
storage memory of the instrument, a warning
appears which you must confirm. Choose Clear
to start the memory clear operation.
Measurement (cont)