624300/01 page 33 / 67
2.7.2 Definitions of Calibration Points
2.7.2.1 Calibration Points 1 and 2 (Standard Calibration)
The limits for the calibration point 1 are defined in register 5152, for calibration point 2 in register 5184.
Start
register
Number of
registers
Reg1 / Reg2 Reg3 / Reg4 Reg5 / Reg6 Modbus
function code
Read
access
Write
access
5152 6 Physical unit
currently
active for
CP1
Min value for
CP1 (in the
physical unit
as defined in
Reg1 and 2)
Max value for
CP1 (in the
physical unit
as defined in
Reg1 and 2)
3, 4 U/A/S none
5184 6 Physical unit
currently
active for
CP2
Min value for
CP2 (in the
physical unit
as defined in
Reg1 and 2)
Max value for
CP2 (in the
physical unit
as defined in
Reg1 and 2)
3, 4 U/A/S none
Figure 2.7.2.1.1: Definition of register 5152 for CP1 and 5184 for CP2.
Attention:
The only physical unit available for calibration is pH ! The physical unit defined in 5152, 5184 and 5312
for CP1, CP2 and CP6 is NOT linked to the physical unit defined for PMC1 in register 2090.
When performing a calibration while having the physical unit set to mV, the calibration status will
report CP1: incorrect measurement unitor CP2: incorrect measurement unit(see figure 2.7.4.1.1).
Command: Calibration limits CP1
Modbus address: 5152 Length: 6 Type: 3
Read
Parameter: Unit Min value Max value
Format: Hex Float Float
Value:
0x01000 0 0
Figure 2.7.2.1.2: Example to read the limits of CP1. Currently active physical unit is pH (0x01000). The
min and max values are both 0, indicating, that calibration at CP1 can be performed only using
defined calibration standards having discrete pH values.
Command: Calibration limits CP2
Modbus address: 5184 Length: 6 Type: 3
Read
Parameter: Unit Min value Max value
Format: Hex Float Float
Value:
0x01000 0 0
Figure 2.7.2.1.3: Example to read the limits of CP2. The active physical unit is is pH (0x01000). The
min and max values are both 0, indicating, that calibration at CP2 can be performed only using
defined calibration standards having discrete pH values.
When initiating the calibration at CP1 and CP2, the measured pH and temperature have to be stable for
at least 3 minutes. The stability criteria are defined in register 5128:
Start
register
Number of
registers
Reg1 / Reg2
(Float)
Reg3 / Reg4
(Float)
Modbus
function code
Read
access
Write
access
5128 4 Max. Drift PMC1
pH
[pH/min]
Max. Drift PMC6
Temperature
[K/min]
3, 4, 16 U/A/S S
Figure 2.7.2.1.4: Definition of register 5128.