2.7.10.2 Definitions for Individual Sets of Calibration Standards
Once the operator has selected a set of calibration standards (register 9474) the register 9504 and
following give all information on the selected set of calibration standards.
Start
register
Number of
registers
Reg1 Reg8
(16 ASCII characters)
Modbus
function code
Read
access
Write
access
9504 8 Manufacturer of the selected set 3, 4 U/A/S none
9512 8 Info 1 of the selected set 3, 4 U/A/S none
9520 8 Info 2 of the selected set 3, 4 U/A/S none
Figure 2.7.10.2.1: Definition of registers 9504 to 9520.
Command: manufacturer of set
Modbus address: 9504 Length: 8 Type: 3
Read
Parameter: Text
Format: Character
Value:
HAMILTON
Figure 2.7.10.2.2: Example to read the description (in this case, the set 0x01 is selected)
Within one set of calibration standards, a maximum of 12 calibration standards are available. In the
following registers, the details of each calibration standard is given:
Start register Description
9536 Nominal value and tolerance of calibration standard 1
9552 Nominal value and tolerance of calibration standard 2
9568 Nominal value and tolerance of calibration standard 3
9584 Nominal value and tolerance of calibration standard 4
9600 Nominal value and tolerance of calibration standard 5
9616 Nominal value and tolerance of calibration standard 6
9632 Nominal value and tolerance of calibration standard 7
9648 Nominal value and tolerance of calibration standard 8
9664 Nominal value and tolerance of calibration standard 9
9680 Nominal value and tolerance of calibration standard 10
9696 Nominal value and tolerance of calibration standard 11
9712 Nominal value and tolerance of calibration standard 12
Figure 2.7.10.2.3: Definition for the register range from 9536 until 9720.
Start
register
Number
of
registers
Reg1 /
Reg2
(Float)
Reg3 /
Reg4
(Float)
Reg5 /
Reg6
(Float)
Reg7 /
Reg8
(Float)
Modbus
function
code
Read
access
Write
access
9536,
9552,
8 Calibr.
standard
nominal
value
[pH]
Tolerance
of nominal
value (±)
[pH]
Nominal
electrical
potential
[mV]
Tolerance
of electrical
potential
(±) [mV]
3, 4 U/A/S none
Figure 2.7.10.2.4: Definition for registers 9536 until 9720.
Command: Read standard 4
Modbus address: 9584 Length: 8 Type: 3
Read
Parameter: Nominal value
[pH]
Tolerance
(±) [pH]
Nominal electrical
potential
[mV]
Tolerance of
electrical potential
(±) [mV]
Format: Float Float Float Float
Value:
4.01 0.02 180 80
Figure 2.7.10.2.5: Example to read the values for the calibration standard 4.