Hamilton Electronics EPHUM011 Stud Sensor User Manual


 
pH ARC Sensors Modbus RTU Programmers Manual (EPHUM011)
624300/01 page 43 / 67
2.7.3.2.4 Product calibration: Restore standard calibration
If a product calibration is active this product calibration can be temporarily switched off by writing the
command 0x03 to register 5340.
Performing this action the values of the product calibration remain stored in the sensors memory.
Command: CP6: Restore standard
Modbus address: 5340 Length: 2 Type: 16
Write
Parameter: Command
Format: Hex
Value:
0x03
Figure 2.7.3.2.4.1: Example to restore a standard calibration from an active product calibration. Writing
command 0x03 (restore standard calibration) to register 5340.
From now on the sensor is measuring using its prior CP1 / CP2 standard calibration.
The sensors calibration status (register 5318) will be reading CP6 assigned(0x10000000) meaning
that a valid assignment for a product calibration is available in the sensors memory (see figure
2.7.4.1.1).
2.7.3.2.5 Product calibration: Restore product calibration
If a valid but inactivated product calibration is available in the sensors memory, the calibration status is
reading CP6 assigned(corresponding to 0x10000000, see figure 2.7.4.1.1), this stored product
calibration can be restored or reactivated by writing command 0x04 to register 5340.
Command: CP6: Restore product
Modbus address: 5340 Length: 2 Type: 16
Write
Parameter: Command
Format: Hex
Value:
0x04
Figure 2.7.3.2.5.1: Example to restore an available product calibration from an active standard
calibration. Writing command 0x04 (restore product calibration) to register 5340.
From now on the sensor is measuring using its prior CP6 product calibration.
The sensors calibration status (register 5318) will be reading 0x14000000 (corresponding to CP6
assignedand CP6 active, see figure 2.7.4.1.1) again.
If this command is performed without available product calibration in the sensors memory the sensor
will respond with a Modbus exception since this command is not valid.