Hamilton Electronics EPHUM011 Stud Sensor User Manual


 
pH ARC Sensors Modbus RTU Programmers Manual (EPHUM011)
624300/01 page 39 / 67
Example to read the sensors quality indicator:
Command: Quality indicator
Modbus address: 4872 Length: 2 Type: 3
Read
Parameter: Quality [%]
Format: Float
Value:
100
Figure 2.7.3.1.9: Example to read the sensors quality indicator
2.7.3.2 Calibration at CP6 (Product Calibration)
The product calibration is a process in order to adjust the measurement of a correctly calibrated pH
ARC Sensor to specific process conditions.
Product calibration is a two stage process:
1. An initial measurement is performed while the operator takes a sample of the process solution.
At that time point the pH ARC Sensor stores its raw measurement value, temperature and
operating hour in the memory.
While the operator takes the sample to the analytics lab for reference analysis the pH ARC
Sensor is still running on its prior standard calibration (CP1 and CP2) while the initial
measurement data for the ongoing product calibration is kept in the sensors memory.
2. When the result of the reference analysis is available this value is assigned, at a second time
point, to the former initial measurement data stored in the pH ARC Sensor.
The sensor is now, after valid assignment, running on a calibration function which is
compensated for the correct process conditions. The product calibration (CP6) is now active.
Performing a Cancel command for the product calibration (CP6) brings the sensor back to its still stored
standard calibration (CP1 and CP2).
If a product calibration is still active and a standard calibration (CP1 or CP2) is performed the product
calibration (CP6) is cancelled.
If the operator needs to overrun an active product calibration (old CP6) by a new product calibration
(new CP6) the above process applies in the same way. After initial measurement the pH ARC Sensor
is still running on the first product calibration (old CP6) until a valid assignment has been done (new
CP6).