LEAK TESTING
This function enables testing for leaks in pressure systems. The
instrument has a function to automate leak testing in a variety of
applications.
Automatic leak test results are logged, not stored and cannot be
output directly via HHP-2000DL into your MLWIN software. To send
leak test results to your PC please use your PC’s Hyper Terminal and
output as to an Epson Printer. When entering pressure values (P1,
P2, P3) you must enter numbers to exactly the same number of
decimal places as full pressure range of the instrument being used.
The number of decimal places required can be seen in the table of
pressure ranges on the back page.
LEAK TESTING
This function enables testing for leaks in pressure systems. The
instrument has a function to automate leak testing in a variety of
applications.
Automatic leak test results are logged, not stored and cannot be
output directly via HHP-2000DL into your MLWIN software. To send
leak test results to your PC please use your PC’s Hyper Terminal and
output as to an Epson Printer. When entering pressure values (P1,
P2, P3) you must enter numbers to exactly the same number of
decimal places as full pressure range of the instrument being used.
The number of decimal places required can be seen in the table of
pressure ranges on the back page.
AVERAGING
MAX/MIN
MAX/MIN
Press key (when in mode) and symbol, together
with average reading, will appear on display. Press
to return to mode.
AVERAGING
MAX/MIN
MAX/MIN
Press key (when in mode) and symbol, together
with average reading, will appear on display. Press
to return to mode.
SETTING UP
Press key (when not in mode) and and
symbols will appear on the display. Enter pre-stress pressure level
(if pre-stress pressure function is not required, press key) and the
display will show and symbols. Key in start pressure and
press key and the and symbols will now be displayed.
Key in test time in minutes and seconds (MM.SS), press key and
the and symbols will now be displayed. Key in permissable
change in pressure and press key. The instrument is now ready
to run test and will appear on display.
MAX/MIN
NE
NE P1
P2
NE
NE P3
P1
SETTING UP
Press key (when not in mode) and and
symbols will appear on the display. Enter pre-stress pressure level
(if pre-stress pressure function is not required, press key) and the
display will show and symbols. Key in start pressure and
press key and the and symbols will now be displayed.
Key in test time in minutes and seconds (MM.SS), press key and
the and symbols will now be displayed. Key in permissable
change in pressure and press key. The instrument is now ready
to run test and will appear on display.
MAX/MIN
NE
NE P1
P2
NE
NE P3
P1
X
X
X
X
X
X
OUTPUT OF DATA
1
2
3
This feature will only function if either the or symbols are
displayed.
To view data press key, then and keys. Thefirst stored
reading will be displayed. Press key to view successively stored
readings and any other key to return to normal mode.
To output data to a PC press key, then key.
Ensure PC/Printer is ready to
receive information.
After transmission, the unit will return to normal operation and any
auto logging will stop.
To erase stored data press key, then and keys.
There will be a delay before the unit will accept the key and
symbol will be displayed. Thisis to help stop accidental erasure.
After erasure, neither or symbols will be displayed. Theunit
will return to normal mode.
The selected transmit option will be displayed during transmission.
If after pressing key you do not wish to
delete data, you must turn instrument off and on, then the stored
data will not be lost.
To output data to an Epson compatible printer press key, followed
by key. Printer settings must be as follows: 9600 Baud, 8 data bits,
No Parity, One stop bit, No flow control. It is also possible to print to
Hyper Terminal in your PC in this way.
4
3
OUTPUT OF DATA
1
2
3
This feature will only function if either the or symbols are
displayed.
To view data press key, then and keys. Thefirst stored
reading will be displayed. Press key to view successively stored
readings and any other key to return to normal mode.
To output data to a PC press key, then key.
Ensure PC/Printer is ready to
receive information.
After transmission, the unit will return to normal operation and any
auto logging will stop.
To erase stored data press key, then and keys.
There will be a delay before the unit will accept the key and
symbol will be displayed. Thisis to help stop accidental erasure.
After erasure, neither or symbols will be displayed. Theunit
will return to normal mode.
The selected transmit option will be displayed during transmission.
If after pressing key you do not wish to
delete data, you must turn instrument off and on, then the stored
data will not be lost.
To output data to an Epson compatible printer press key, followed
by key. Printer settings must be as follows: 9600 Baud, 8 data bits,
No Parity, One stop bit, No flow control. It is also possible to print to
Hyper Terminal in your PC in this way.
4
3
Press key. symbol and will appear on the display.
Enter required time interval in hh:mm (ie, in order to take a reading
every 1 minute, key in 0001) and press key. Current reading will
return to display and symbol will remain on display.
NE 0000
Instrument
will continue logging until a maximum of 250 readings are taken or
data is down-loaded.
Press key. symbol and will appear on the display.
Enter required time interval in hh:mm (ie, in order to take a reading
every 1 minute, key in 0001) and press key. Current reading will
return to display and symbol will remain on display.
NE 0000
Instrument
will continue logging until a maximum of 250 readings are taken or
data is down-loaded.
WARRANTY / DISCLAIMER
RETURN REQUESTS / INQUIRIES
OMEGAENGINEERING,INC warrants this unit to be free of defects in materials and
workmanship for a period of 25 months from the date of purchase. OMEGAWarranty
adds an additional one (1) month grace period to the normal one (1) year product
warranty to cover handling and shipping time. Thisensures that OMEGA’S
customers receive maximum coverage on each product. If the unit malfunctions, it
must be returned to the factory for evaluation. OMEGA’s Customer Service Depart-
ment will issue anAuthorizedReturn (AR) number immediately upon phone or written
request. Upon examination by OMEGA, if the unit is found to be defective, it will be
repaired or replaced at no charge. OMEGA’S WARRANTY does not apply to defects
resulting from any action of the purchaser, including but not limited to mishandling,
improper interfacing, operation outside of design limits, improper repair, or
unauthorized modification. ThisWARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of
excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of OMEGA’s control.
Components which wear are not warranted, including but not limited to contact points,
fuses, and triacs. OMEGAis pleased to offer suggestions on the use of its various
products. However, OMEGAneither assumes responsibility for any omissions or
errors nor assumes liability for any damages that result from the use if its products in
accordance with information provided by OMEGA, either verbal or written. OMEGA
warrants only that the parts manufactured by it will be as specified and free of
defects. OMEGAMAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE,
ANDALLIMPLIED WARRANTIES INCLUDINGANY WARRANTY OF MERCHANT-
ABILITYAND FITNESS FOR APARTICULAR PURPOSE ARE HEREBY DIS-
CLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein
are exclusive, and the total liability of OMEGAwithrespectto this order, whether
based on contract, warranty, negligence, indemnification, strict liability or otherwise,
shall not exceed the purchase price of the component upon which liability is based.
In no event shall OMEGAbeliable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGAisnot intended to be used, nor shall it be
used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any
nuclear installation or activity; or (2) in medical applications or used on humans.
Should any Product(s) be used in or with any nuclear installation or activity, medical
application, used on humans, or misused in any way, OMEGA assumes no
responsibility as set forth in our basic WARRANTY/DISCLAIMERlanguage, and,
additionally, purchaser will indemnify OMEGA and hold OMEGAharmless from any
liability or damage whatsoever arising out of the use of the Product(s) in such a
manner.
Direct all warranty and repair requests/inquiries to the OMEGACustomerService
Department. BEFORE RETURNINGANY PRODUCT(S) TO OMEGA, PURCHASER
MUST OBTAIN ANAUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S
CUSTOMER SERVICE DEPARTMENT (IN ORDERTO AVOID PROCESSING
DELAYS). The assignedAR number should then be marked on the outside of the
return package an on any correspondence.
FOR WARRANTYRETURNS, ForNON-WARRANTY REPAIRS,
please have the following information consult OMEGAfor current repair
available BEFORE contacting OMEGA: charges. Have the following information
1. Purchase Order number under which available BEFORE contacting OMEGA:
the product was PURCHASED, 1. Purchase Order number to cover the
2. Model and serial number of the COST of the repair/calibration
product under warranty, and 2. Model and serial number of the
3. Repair instructions and/or specific product, and
problems relative to 3. Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGAENGINEERING, INC.
© Copyright 1998 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in
whole or in part, without the prior written consent of OMEGA ENGINEERING, INC. It is the policy of
OMEGA to comply with all worldwide safety and EMC / EMI regulations that apply. OMEGA is constantly
pursuing certification of its products to the European New Approach Directive. OMEGA will add the CE
mark to every appropriate device upon certification. The information contained in this document is believed
to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the
right to alter the specifications without notice. WARNING: These products are not designed for use in, and
should not be used for, patient connected applications.
WARRANTY / DISCLAIMER
RETURN REQUESTS / INQUIRIES
OMEGAENGINEERING,INC warrants this unit to be free of defects in materials and
workmanship for a period of 25 months from the date of purchase. OMEGAWarranty
adds an additional one (1) month grace period to the normal one (1) year product
warranty to cover handling and shipping time. Thisensures that OMEGA’S
customers receive maximum coverage on each product. If the unit malfunctions, it
must be returned to the factory for evaluation. OMEGA’s Customer Service Depart-
ment will issue anAuthorizedReturn (AR) number immediately upon phone or written
request. Upon examination by OMEGA, if the unit is found to be defective, it will be
repaired or replaced at no charge. OMEGA’S WARRANTY does not apply to defects
resulting from any action of the purchaser, including but not limited to mishandling,
improper interfacing, operation outside of design limits, improper repair, or
unauthorized modification. ThisWARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of
excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of OMEGA’s control.
Components which wear are not warranted, including but not limited to contact points,
fuses, and triacs. OMEGAis pleased to offer suggestions on the use of its various
products. However, OMEGAneither assumes responsibility for any omissions or
errors nor assumes liability for any damages that result from the use if its products in
accordance with information provided by OMEGA, either verbal or written. OMEGA
warrants only that the parts manufactured by it will be as specified and free of
defects. OMEGAMAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE,
ANDALLIMPLIED WARRANTIES INCLUDINGANY WARRANTY OF MERCHANT-
ABILITYAND FITNESS FOR APARTICULAR PURPOSE ARE HEREBY DIS-
CLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein
are exclusive, and the total liability of OMEGAwithrespectto this order, whether
based on contract, warranty, negligence, indemnification, strict liability or otherwise,
shall not exceed the purchase price of the component upon which liability is based.
In no event shall OMEGAbeliable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGAisnot intended to be used, nor shall it be
used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any
nuclear installation or activity; or (2) in medical applications or used on humans.
Should any Product(s) be used in or with any nuclear installation or activity, medical
application, used on humans, or misused in any way, OMEGA assumes no
responsibility as set forth in our basic WARRANTY/DISCLAIMERlanguage, and,
additionally, purchaser will indemnify OMEGA and hold OMEGAharmless from any
liability or damage whatsoever arising out of the use of the Product(s) in such a
manner.
Direct all warranty and repair requests/inquiries to the OMEGACustomerService
Department. BEFORE RETURNINGANY PRODUCT(S) TO OMEGA, PURCHASER
MUST OBTAIN ANAUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S
CUSTOMER SERVICE DEPARTMENT (IN ORDERTO AVOID PROCESSING
DELAYS). The assignedAR number should then be marked on the outside of the
return package an on any correspondence.
FOR WARRANTYRETURNS, ForNON-WARRANTY REPAIRS,
please have the following information consult OMEGAfor current repair
available BEFORE contacting OMEGA: charges. Have the following information
1. Purchase Order number under which available BEFORE contacting OMEGA:
the product was PURCHASED, 1. Purchase Order number to cover the
2. Model and serial number of the COST of the repair/calibration
product under warranty, and 2. Model and serial number of the
3. Repair instructions and/or specific product, and
problems relative to 3. Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGAENGINEERING, INC.
© Copyright 1998 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in
whole or in part, without the prior written consent of OMEGA ENGINEERING, INC. It is the policy of
OMEGA to comply with all worldwide safety and EMC / EMI regulations that apply. OMEGA is constantly
pursuing certification of its products to the European New Approach Directive. OMEGA will add the CE
mark to every appropriate device upon certification. The information contained in this document is believed
to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the
right to alter the specifications without notice. WARNING: These products are not designed for use in, and
should not be used for, patient connected applications.