A SERVICE OF

logo

MG500-EU-EN v1.0 7/13
7
Operation
Poweringthemeter
PresstheONbuttontoturntheuniton.Pressthe OFFbuttontoturntheunitoff.Ifthemeter
doesnotswitchONpleasecheckthatfresh‘C’1.5Valkalinebatteriesareinstalledinthebattery
compartmentlocatedatthebottomoftheinstrument.
InsulationResistanceTestingProcedure
Warning:Thecircuitundertestmustbecompletelydeenergizedandisolatedbeforemaking
testconnections.
1. Afterturningtheuniton,makesurethattheReplaceBatterymessageisnotdisplayed.If
theReplaceBatterymessageisdisplayedreplacethebatteriesbeforeproceeding.
2. Thetestvoltagedefaultsto
2500Vonpowerup.Selectthedesiredtestvoltageusingthe
Voltage Adjust buttons (±500V) that step through the available test voltages in 500V
incrementsorbypressingoneofthepresettestvoltagebuttons(10kV,5kV,2.5kV,1Kv).
3. ThedisplayshowstheConnectLeadspromptalongwiththeselected
testvoltagevalue.
ConnecttheGreentestleadtotheGreentestsocketonthemeterandthentotheGround
sideoftheinsulatorundertest.
ConnecttheRedleadtotheRedtestsocketonthemeterandtheBlueguardcabletothe
bluesocketonthe
meter.ConnecttheserviceendoftheRedcabletoothersideofthe
insulatorundertest.
IfaLeakageGuardleadisrequiretocollectsurfaceleakage,connecttheseparateBluetest
leadinto theBlue guardleadfromthe Redtest leadwhich ispluggedintothe Bluetest
socketonthemeterandthencliptheserviceendofthecabletothesurfaceoftheinsulator
undertest.
4. MomentarilyPresstheTESTSTART/STOPbuttontostarttesting.
ForDARorPItestingrefertotheseparatetestprocedurebelow.
5. The Insulationresistance test will
completein 10 seconds.The test voltagepercentage
steadilycountsdowntozeroduringthetestascanbeseenonthedisplay.Theresistance
testresultvaluewillappearonthedisplaywhenthetesthascompleted.
6. TheHOLDiconwillappearinplaceofthebargraphwhenthe
testiscompleteandafterthe
voltagehasbeendischarged.
WARNING:DONOTREMOVETHETESTLEADSBEFORETHEHOLDDISPLAYAPPEARS.
7. Themeterautomaticallydischargesthesystemattheendofthetest.
8. Totakeanothertest,presstheTESTSTART/STOPbutton,theConnectLeadspromptshould
appearonthedisplay;theinstrumentisnowreadytorunanewtest.
9. Theinstrumentautomaticallypowersdownafter5minutesofinactivity.