MG500-EU-EN v1.0 7/13
9
BargraphVoltageDisplay
Thebargraphrepresentsthevoltagepresentonthetestleadsasitrises,soaks,anddecays.The
bargraphappearsonthelowerleftofthedisplaywindowduringatest.
AutomaticUnder/OverRangeResistanceDetect
Ifthedisplayshowsthemessage“LOWM‐ohms”,thetestshouldbeinterruptedimmediatelyby
pressingtheTESTbutton.Thismessageindicatesthattheinsulationundertesthasbrokendown
andthemeteristryingtoinjectahighpotentialontoashortcircuit.If‘HIGH’appears,itislikely
thatthetestconfigurationisopen.Checkthetestset‐upandthentestagain.
ElapsedTimer
Thetestdurationisindicatedonthedisplay.Thisisparticularlyusefulindeterminingwhetheran
insulatingmaterialundertestwillbreakdowninagivenamountoftime.
ManualTestSTOP
Tostopatestinprogress,presstheTESTbutton.Thetestwillimmediatelyendandthesystem
willautomaticallydischarge.
AutomaticTestSTOP
WheninthedefaultEnersave
TM
mode,thetestautomaticallystopsafterapprox.10seconds.
WhennotintheEnersave
TM
modethetestautomaticallystopsafterapprox.100seconds.
LiveCircuitWarning
Ifthetestleadsareconnectedtoalivecircuit(approx.500V),awarningbeeperwillsoundand
themeterwilldisplay“LiveWarning…CircuitLive…”Inthiscase,correcttheproblemandretest.
AutomaticDischarge
At the end of a test, the meter automatically discharges the high voltage. The automatic
dischargestatusisreflectedonthedisplay.Duringdischargethebeeperwillsoundand,when
completelydischarged,theHOLDiconwillappearonthedisplay.