46
Retrotec Energy Innovations, Ltd.
DM-2A Mark II 2008-02-22
3. Press the enteR keypad button to display the Enable/Disable Device screen
showing a list of device profile names. A Yes status in the right-hand column
indicates the device profile is enabled; a No indicates the profile is currently
disabled.
4. Select your device and toggle its enable/disable state using the (Left), and/or
(Right) keypad buttons.
5. Press the exit keypad button to return to the main Setup menu and again to
return to the main operational screen.
6. For more information on enabling or disabling test devices, see Choosing Mode
Setup Options in Section 2 Working with the Setup Screen.
Q: The room pressure value my gauge is displaying doesn’t appear to be correct? What’s
causing this?
A: This is often caused by the Baseline mode inadvertently being left on. Check the
status information next to the Baseline label on the display. To toggle the Baseline
status off, press the BaSeline keypad button once.
Q: The pressure values on my gauge display intermittently and the gauge itself seems
to be operating sluggishly. Is it malfunctioning?
A: It may be that the Time Average feature is active and set to a long time interval.
Check the status information next to the Time label on the display. To turn the Time
Average status off or decrease the interval time period, press the tiMe avg keypad
button to cycle through the available options.
Q: I have tried to resolve my problem on my own, but I believe my gauge is
malfunctioning. What do I do?
A: If you are unable to resolve your current problem on your own, contact technical
support using the contact information in the following section.
DM-2A Mark II Dual-Channel Pressure Gauge User Manual
C: Getting Help