SETTING INSTRUMENT CONFIGURATION
3-5
Tab Down Tab Up Enter Back Read On/Off
Exiting Communication Option
• After you have completed all the communications editing, press Back to return to
the Configuration menu.
Languages
The Language configuration allows you to select the language you want to display on
your instrument.
1. Press Tab Up or Tab Down to move the highlight to Language.
CONFIGURATION
Display Options
Measure Options
Set Clock
Communications
Language
2. Press Enter to access the Language editor.
Language
English Chinese Simp
Deutsch Japanese
Français
Espanol
Italiano
Portugues
3. Press Tab Up or Tab Down to move the highlight to the desired language.
4. Press Enter to save the selected language. The instrument returns to the Configuration
menu with the selected language active.
Battery Refresh Operation
Your instrument is powered by a Ni-metal hydride battery pack. Because the battery will
start to charge every time the instrument is docked in the station, this can, overtime reduce
the capacity of the battery. Refreshing is the process of allowing your instrument’s battery
to drain completely and then charging all cells in the battery to 100%. Refreshing the
battery from time-to-time helps extend the duration of use and life of the battery.
The process for refreshing the battery takes approximately eight hours and should be
performed once a month. As the instrument cannot be used during this operation and
must remain in the docking station the entire time, it is suggested that this procedure be
performed at the end of a workday to minimize disruption.
1. Place the instrument in the docking station.
2. Press Tab Up or Tab Down to move the highlight to the Battery icon and then press Enter.
CONFIGURATION
Display Options
Measure Options
Set Clock
Communications
Language
3. You are asked if you are sure you would like to begin the refresh operation. Press Tab
Up or Tab Down to move the highlight to Yes and press Enter.
The instrument displays “Please Wait, battery maintenance in progress” during the
discharge phase of the operation. This screen automatically disappears when the charging
portion of the operation begins.
NOTE: To abort the refresh operation at anytime, simply power down the
instrument and then power it back up.