Agilent Technologies E3632A Welding System User Manual


 
5
Display Annunciators
Adrs Power supply is addressed to listen or talk over a remote interface.
Rmt Power supply is in remote interface mode.
15V Shows the 15V/7A range is selected.
30V Shows the 30V/4A range is selected.
OVP The overvoltage protection function is enabled when the annunciator
turns on or the overvoltage protection circuit has caused the power
supply to shutdown when the annunciator blinks.
OCP The overcurrent protection function is enabled when the annunciator
turns on or the overcurrent protection circuit has caused the power
supply to shutdown when the annunciator blinks.
CAL The power supply is in calibration mode.
Limit The display shows the limit values of voltage and current.
ERROR Hardware or remote interface command errors are detected and the error
bit has not been cleared.
OFF The output of the power supply is disabled (See
page 52 for more
information
).
Unreg The output of the power supply is unregulated (output is neither CV
nor CC).
CV The power supply is in constant voltage mode.
CC The power supply is in constant current mode.
To review the display annunciators, hold down key as you
turn on the power supply.
Display Limit