Agilent Technologies E3632A Welding System User Manual


 
Chapter 4 Remote Interface Reference
Simplified Programming Overview
80
3RZHU6XSSO\3URJUDPPLQJ5DQJHV
7KH
SOURce
VXEV\VWHPUH TXLUHVSD UD PH WHU VIRUSU RJU D PPLQ JY D OXH V7KH
DYDLOD EOHSURJU DP PLQ JY DOX HIRU DSD UD PH WHU YD ULHVDFF RU GLQJ WRWKHG HVLU HG 
RXWSXWUDQJH RIWKHSRZHUVXSSO \7KHIROORZLQJWDE OHOL VWVWKHSUR JU DP PLQ J
YDOXHVD YDLODEOH DQ G0,1LP XP0$ ; LPXP '()D XOWDQG UH VHWYD OX HVRIWKH
$JLOLHQW( $ SRZH UVXSS O\
Refer to this table to identify programming values when programming the power
supply.
Table 4-1
. Agilent E3632A Programming Ranges
0 - 15V/7A Range 0 - 30V/4A Range
Voltage Programming Range 0 V to 15.45V 0 V to 30.9 V
MAX Value 15.45 V 30.9 V
MIN Value 0 V 0 V
DEFault Value 0 V 0 V
*RST Value 0 V
Current Programming Range 0 A to 7.21 A 0 A to 4.12 A
MAX Value 7.21 A 4.12 A
MIN Value 0 A 0 A
DEFault Value 7 A 4 A
*RST 7 A