474 Chapter5
One-Button Measurement Functions
Meas Setup (Spectrum Emissions Mask—SEM)
One-Button Measurement Functions
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST
ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative,ABSolute|AND|OR|RELative
,ABSolute|AND|OR|RELative
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST?
Remote Command Notes: Comma separated list of 5 values. n = 1 for BTS and n = 2 for
MS. Default is BTS (1).
Example: SEM:OFFS:LIST:TEST ABS,REL,ABS AND REL,ABS OR REL
SEM:OFFS:LIST:TEST?
5.22.4.11.1 Absolute
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement
resultsislargerthanthelimitfor
Abs Start and/or Abs Stop. This is the default selection for
each offset
Key Path:
Meas Setup, Offset/Limits, More, Fail Mask
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for
“Offset/Limits” on page 462 for this value.
Remote Command:
Refer to
“Fail Mask” on page 472.
5.22.4.11.2 Relative
Sets the limit test to show Fail if one of the relative spectrum emission mask measurement
resultsislargerthanthelimitfor
Rel Start and/or Rel Stop.
Key Path:
Meas Setup, Offset/Limits, More, Fail Mask
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for
“Offset/Limits” on page 462 for this value.
Remote Command:
Refer to
“Fail Mask” on page 472.
5.22.4.11.3 Abs AND Rel
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement
resultsislargerthanthelimitfor
Abs Start and Abs Stop AND one of the relative spectrum
emission mask measurement results is larger than the limit for
Rel Start and Rel Stop.