MAINTENANCE
Generally, the only maintenance required is to keep the opening of the sensor clear of deposits. Cleaning
frequency is best determined by experience.
MODEL 228 MAINTENANCE
WARNING
Retraction chamber contents may be under pressure.
CAUTION
Be sure sensor has been cleaned of process liquid
before handling.
TROUBLESHOOTING
PROBLEM PROBABLE CAUSE SOLUTION
Off-scale reading Wiring is wrong. Verify and correct wiring.
RTD is open or shorted. Check the RTD for open or short circuits.
See Figure 19.
Sensor is not in process stream. Confirm that the sensor is fully submerged in
the process stream. See installation section.
Sensor is damaged. Perform isolation checks. See Figure 19.
Noisy reading Sensor is improperly installed in the Confirm that the sensor is fully submerged
process stream. in the process stream. See installation section.
Sensor cable is run near high voltage Move cable away from high voltage conductors.
conductors.
Sensor cable is moving. Keep sensor cable stationary.
Reading seems wrong (lower Bubbles are trapped in the sensor, Install the sensor in a vertical pipe run with the
or higher than expected) particularly in the toroid opening. flow against the toroid opening. Increase flow
if possible.
Sensor is not completely submerged in the Confirm that the sensor is fully submerged
process stream. in the process stream. See installation section.
Cell constant is wrong. Wall effects are Calibrate the sensor in place in the process
present. piping. See calibration section.
Wrong temperature correction algorithm Check that the temperature correction is
is being used. appropriate for the sample. See analyzer
manual for more information.
Temperature reading is inaccurate. Disconnect the RTD leads (Figure 19) and
measure the resistance between the in and
common leads. Resistance should be close
to the value in Table 1.
Slow temperature response to sudden Use an RTD in a metal thermowell for
changes in temperature. temperature compensation.
Sluggish response Sensor is installed in a dead area in the Move sensor to a location more representative
piping. of the process liquid.
Slow temperature response to sudden Use an RTD in a metal thermowell for
changes in temperature. temperature compensation.
14