MODEL 228 INSTALLATION
INSTALLATION
Keep at least 1 in. (2.5 cm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensor
in place. Ensure that the sensor is completely submerged in liquid. Mounting the sensor in a vertical pipe run with
the flow from bottom to top is best. If the sensor must be mounted in a horizontal pipe run, orient the sensor in the
3 o’clock or 9 o’clock position.
WIRING
Keep sensor wiring away from ac conductors and high current demanding equipment. Do not cut cable.
Cutting the cable may void the warranty.