Output(s) value on analyzer failure:
Choose the desired signal level to cause a failure condi-
tion. The choices are:
Acutal
BeginOfRange
EndOfRange
BeginOfRange-10%
BeginOfRange+10%
Output number:
Choose the output number (1-8) for setting the fine ad-
justment.
Operation mode:
Normal: The absolute measurement signal will be sent
to the analog output.
Adjust 0V: Used to set the display equal to the analog
output for 0 V and 0 mA. Life zero signals (4 - 20 mA and
2 - 10 V) are set automatically and cannot be adjusted.
Adjust 10V: Used to set the display equal to the analog
output for 10V und 20mA.
Select the "Fine adjustment for 0% output" and/or "Fine
adjustment for 100 % output" lines with the or softkey.
Adjust to the desired value with the orsoftkey and
confirm with the softkey.
The range of values are:
3,000 to 6,000 for 0 % (default 4096)
600 to 1,000 for 100 % (default 819)
The last three lines of the "Analog Output Setup" menu
are displayed only for configuration values of the analog
output.
Signal name: The name of the signal chosen the
"Choose signal" menu.
Current signal value: The current value of the variable.
Source module: The name of the module chosen in
the "Choose signal source module" menu.
Pressing the F5 (More...) softkey changes to the
submenu "Special Scaling for Concentration Signal"
Figure 3-16: Special Scaling for Concentration
Signal Menu
This menu allows for the setting of each of the 8 outputs
to be the same as the range limits "Yes" or as to set on
the previous menus.
Pressing the F5 (More...) softkey changes to the
submenu "Analog Output Updates per Second."
Figure 3-17: Analog Output Updates per
Second Menu.
This menu allows for the setting of the update rate for
each of the 8 outputs.
Line#1
Measure More...
0.000 ppm
Back...
-- Special Scaling for Concentration Signal --
(Scaling is the same as range limits)
Output #1:
Output #2:
Output #3:
Output #4:
Output #5:
Output #6:
Output #7:
Output #8:
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Line#1
Measure
0.000 ppm
Back...
-- Analog Output Updates Per Second --
Output #1:
Output #2:
Output #3:
Output #4:
Output #5:
Output #6:
Output #7:
Output #8:
1
0
0
0
0
0
0
0
0