Emerson HAS60E-IM-HW Stud Sensor User Manual


 
NGA 2000 CLD
Instruction Manual
HAS60E-IM-HW
June 2007
4 - 8 Maintenance and Service Emerson Process Management GmbH & Co.OHG
If there is a leakage, then proceed as follows:
a) Using Helium (He) for test medium
Supply Analyzer Module with the test medium
with a pressure of approx. 2,000 hPa (15 psig).
Look for the leak with a helium leak detector.
b) Using Nitrogen (N
2
) for test medium
Supply Analyzer Module with the test medium
with a pressure of approx. 2,000 hPa (15 psig).
Liberally cover all fittings, seals, and other pos-
sible sources of leakage with a suitable leak
test liquid such as SNOOP. Bubbling or foam-
ing indicates leakage.
Remove the source of leak.
Perform a leakage test once more as decribed
obove.
.