Emerson IP2046IM Laser Level User Manual


 
4
IP2046/IM
Nov 2006
2.2 Pressure Equipment Directive
The MSP400RH transmitter does not fall within the PED definition as enclosing a pressurised fluid, so is
therefore outside the scope of the Directive.
Accordingly, the Declaration of Conformity does not list the Pressure Equipment Directive.
2.3 Specifications
Materials of construction:
Transducer material PVDF
Body & cover Material Glass filled nylon
Cable gland Nylon with Nitrile cable seal
Cover seal Silicone rubber
Cover screws 316 Stainless Steel
Transducer bodyseal EPDM
Electrical
Supply voltage 12-40V dc
Output 4-20mA (3.8 - 20.5mA linear)
Relays (2) : SPST rated 1A at 24Vdc
Communications HART Digital communications (Rev. 5)
Earthing None required
Recommended Cable Two core each 0.22 mm
2
min, shielded
Max. cable length 3000m
Cable resistance 0.1 Ohm per metre length max.
Cable gland Suitable for cable sizes 4mm - 8mm diameter
Operating
Range 0.45 to 11m
Temperature
Ambient -40°C to +70°C
Wetside -30°C to +70°C
Pressure -0.25 bar to 3.0 bar
Weatherproofing IP66/IP67
Certification
The MSP400 range of products are Factory Mutual (Fm) certified as Process Control Equipment meeting
basic electrical, mechanical and fire protection requirements (IP66/IP67).