The pH calibration slope (the GLP slope is the average of the calibration
slopes; the percentage is referred to the ideal value of 59.16 mV/pH).
The pH calibration buffers in calibrating order.
The first pH calibration buffer:
The second pH calibration buffer:
The third pH calibration buffer:
Note: The “OLd” message displayed beside the pH value means that
this buffer was not used during last calibration. Press and hold
down SET/CLR if you want to see calibration date (or time if old
calibration was made in the same day with current calibration).
If “no bUF” message appears on the LCD, the instruments inform
you that calibration was performed in less than three points.
pH resolution during calibration.
Calibration Alarm Time-Out status:
if disabled.
CAL pH