25
SERIAL COMMUNICATION COMMAND PREFIX SET
Press CAL when the command prefix is displayed. The command prefix (0 to
47) will start blinking.
Press the ARROW keys to change the command prefix.
Press CFM to save the modified command prefix value.
Press CAL to escape without saving.
Note: • See the PC interface section on page 29 for a complete explanation.
• The command prefix must be 16 if HI 92000 PC software is used.
CALIBRATION ALARM TIME-OUT SET
Press CAL when the calibration time-out is displayed. Calibration time-out
(OFF or 1 to 7 days) will start blinking.
Press the ARROW keys to change the calibration time-out value.
Press CFM to save the modified calibration time-out value.
Press CAL to escape without saving.
INSTRUMENT ID PARAMETER SET
Press CAL when the instrument ID is displayed. The instrument ID (0000 to
9999) will start blinking.
Press the ARROW keys to change the instrument ID value.
Press CFM to save the modified instrument ID value.
Press CAL to escape without saving.
Note: The instrument ID is downloaded to a PC as part of a logged data, set
to identify its origin.