23
''VSET 0.1; ISET 120'' (6031A)
''VSET 0.09; ISET 50" (6032A)
''VSET 1.0; ISET 5'' (6035A)
d. The DVM reading should be in the range:
0.354 to 0.645Vdc (6030A)
0.085 to 0.115Vdc (6031A)
0.050 to 0.130Vdc (6032A)
0.598 to 1.400Vdc (6035A)
Note the reading.
e. Enter and run the following program:
10 OUTPUT 705; "VOUT?"
20 ENTER 705;A
30 DISP A
40 GOTO 10
50 END
f. The value displayed by the controller should be the value noted in step d:
± 0.080Vdc (6030A)
± 0.007Vdc (6031A)
± 0.020 Vdc (6032A)
± 0.205 Vdc (6035A)
g. Send string:
"VSET 200; ISET 17" (6030A)
''VSET 20; ISET 120'' (6031A)
''VSET 60; ISET 50" (6032A)
''VSET 500; ISET 5" (6035A)
h. The DVM reading should be in the range:
199.785 to 200.215Vdc (6030A)
19.978 to 20.022Vdc (6031A)
59.939 to 60.061 Vdc (6032A)
498.350 to 501.650 Vdc (6035A)
Note the reading.
i. Run the program listed in step e. The value displayed by the controller should be the value noted in step h:
± 0.240Vdc (6030A)
± 0.023 Vdc (6031A)
± 0.068 Vdc (6032A)
± 2.700 Vdc (6035A)
Load Effect (Load Regulation). Constant-voltage load effect is the change in dc output voltage (Eo) resulting from a
load-resistance change from open-circuit to full-load. Full-load is the resistance which draws the maximum rated output
current at voltage Eo. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-5. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to maximum.
b. Turn the unit's power on, and, using DISPLAY SETTINGS pushbutton switch, turn up current setting to full output.
c. Turn up output voltage to:
60Vdc (6030A)
7.0Vdc (6031A)
20.0Vdc (6032A)
200Vdc (6035A)
as read on the digital voltmeter.