Emerson 381 Stud Sensor User Manual


 
10
FIGURE 2-6. Retro-Fit Kit Code 04 Flow Powered Cleaner, Model 381
MODEL 381pH/ORP SECTION 2.0
INSTALLATION
PRESSURE DROP AND RECOMMENDED FLOW
RATES- FLOW POWERED CLEANER (CODE 04)
kPa
240
220
200
180
160
140
120
100
20
15
10
5
POSSIBLE ELECTRODE DAMAGE
ABOVE THE THIS POINT
PRESSURE DROP
RECOMMEND-
ED OPERATING
RANGER
INEFFECTIVE CLEANING BELOW
THIS POINT
12345678
51015202530
gpm @ 60°
liters/min
PSIG
FLOW RATE