After the serial poll is completed, the RQS bit in the Status Byte Register of
the HP 16500C Logic Analysis System will be reset if it was set. Once a bit in
the Status Byte Register is set, it will remain set until the status is cleared
with a *CLS command, or the instrument is reset.
Parallel Poll
Parallel poll is a controller-initiated operation which is used to obtain
information from several devices simultaneously. When a controller initiates
a Parallel Poll, each device returns a Status Bit via one of the DIO data lines.
Device DIO assignments are made by the controller using the PPC (Parallel
Poll Configure) sequence. Devices respond either individually, each on a
separate DIO line; collectively on a single DIO line; or any combination of
these two ways. When responding collectively, the result is a logical AND
(True High) or logical OR (True Low) of the groups of status bits.
Figure 7-3 shows the Parallel Poll Data Structure. The summary bit is sent in
response to a Parallel Poll. This summary bit is the "IST" (individual status)
local message.
The Parallel Poll Enable Register determines which events are summarized in
the IST. The *PRE command is used to write to the enable register and the
*PRE? query is used to read the register. The *IST? query can be used to
read the IST without doing a parallel poll.
Status Reporting
Parallel Poll
7–9