SECTION 3—INSTALLING THE CONSERVING REGULATOR
Pneumatic Oxygen Conserver 16 Part No 1160840
SECTION 3—INSTALLING THE
CONSERVING REGULATOR
ƽ WARNING
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththeregulatorgasketorcylinderpostvalve.
Thesesubstancesmaybecomeflammableinthepresenceof
oxygenandcauseinjury.Avoidtouchingthecylinderpost
valvewheneverpossible.Exampleofitemstoavoidarehand
creamsorlotions,cookingoils,suntanoilsorsunblocksand
similaroilyproducts.Ifyouusethesetypesofproducts,wash
yourhandspriortouse.Ifthecylinderpostvalveshould
comeincontactwithsuchsubstancesitshouldbecleanedoff
withadampclothpriortoconnectingittotheconserving
regulator.
DONOTdirectflowofoxygenatanypersonorflammable
materialwhenopeningthecylinder.
DONOTusewithoutCGA870SealWasher[Metalwasher
withelastomerseal].
DONOTuseplasticyokeseals[Plasticcompressiongasket].
UseCGA870SealWashermadeofViton
®
andbrass.
TheCGA870SealWashersuppliedwiththeoxygen
conservingregulatorisreusable.
CAUTION
DONOTusetoolstotightenT‐handle,thiscanleadtoover
tighteningandwillcausedamagetotheoxygenconserving
regulator.
Ensureallconnectionsaretightandleakfree.
DONOTuseliquidleakdetectortotestforleaks.