SECTION 5—REMOVING THE CONSERVING REGULATOR
Pneumatic Oxygen Conserver 22 Part No 1160840
SECTION 5—REMOVING THE
CONSERVING REGULATOR
ƽ DANGER
NEVERattempttoremovetheoxygenconservingregulator
fromanoxygencylinderunlessthecylindervalveis
completelyclosed.
CAUTION
DONOTtrytoloosentheThandlebeyondthestop.
Otherwise,damagetotheoxygenconservingregulatorwill
occur.
NOTE:Forinformationonhowtooperatethecylinder,consult
yourequipmentprovider.
1. Completelyclosetheoxygencylindervalve.
2. Turntheoxygenconservingregulatorselectorswitchto2
L/min.continuousflowposition.
3. Waitforoxygentostopflowingfromtheoxygen
conservingregulator.
4. TurntheThandletoloosentheconnectiontothepost
valve.
5. Removetheoxygenconservingregulatorfromthepost
valve.