SECTION 8—TROUBLESHOOTING
Part No 1160840 27 Pneumatic Oxygen Conserver
SECTION 8—TROUBLESHOOTING
Ifoxygenconservingregulatorfailstofunction,consultthe
Troubleshootingguide.Ifproblemcannotbecorrected,
consultyourDealer/HomeCareequipmentprovider.
Replacement Parts - Accessories
HF2POST6BAGCarryingBagorHF2POST9BAGCarrying
Bag
PROBLEM POSSIBLE CAUSE SOLUTION
No flow. Unit does not
pulse.
1. Cylinder valve is
closed.
2. Regulator in “Off”
position.
3. Cylinder empty.
4. Oxygen conserving
regulator not
sensing breath.
5. Oxygen cannula is
blocked or kinked.
1. Turn the cylinder
valve to the “ON”
position.
2. Set to prescribed
setting.
3. Replace cylinder.
4. Check position of
cannula in nose. DO
NOT breathe
through mouth.
5. Remove kinks. Clean
or replace, if
necessary.
Unit flows or pulses
continuously.
1. Unit is set to the
“CF” position.
2. Unit was not set to
“OFF” prior to
opening the
cylinder valve.
3. Vent hole is
obstructed.
1. Turn the selector
switch to the
appropriate delivery
setting.
2. Turn the selector
switch to “OFF,” wait
a few moments, then
set at proper delivery
setting.
3. Remove obstructions,
such as labels or
carrying bag, and
resume use as usual.