SECTION 6—TYPICAL PRODUCT PARAMETERS
Part No 1160840 25 Pneumatic Oxygen Conserver
Oxygen Cylinder Duration
Thetotaldeliveryofoxygenviatheconserverisrelatedtothe
breathingrates,itisuseradaptiveinthatthetotaloxygen
deliveredperminutewillautomaticallyadjustwithuser
need,asexpressedbyincreasedordecreasedbreathingrates.
Forexample,atallsettings,twiceasmuchoxygenperminute
willbedeliveredifonebreathestwenty(20)timesperminute
ascomparedwithten(10)timesperminute.Referto
OXYGENDURATIONTABLEonpage 25.
OXYGEN DURATION TABLE
SETTING123456Continuous
Flow
2 L/min.
Cylinder Type Cylinder
Volume
Liters @
2015 psi
Estimated Cylinder Duration in Hours
(Based on 20 breaths/min.)
M2 36 1.9.9.6.5.4.3 .3
M4(A) 113 5.9 2.9 2.0 1.5 1.2 1.0 .9
M6(B) 164 8.5 4.3 2.8 2.1 1.7 1.4 1.4
ML6 171 8.9 4.5 3.0 2.2 1.8 1.5 1.4
M9(C) 246 12.8 6.4 4.3 3.2 2.6 2.1 2.1
D 425 22.1 11.1 7.4 5.5 4.4 3.7 3.5
E 680 35.4 17.7 11.8 8.9 7.1 5.9 5.7