Invacare IOC100P Staple Gun User Manual


 
SECTION 4—OPERATING INSTRUCTIONS
Pneumatic Oxygen Conserver 18 Part No 1160840
SECTION 4—OPERATING
INSTRUCTIONS
DANGER
NEVERsmokeinanareawhereoxygenisbeing
administered.
NEVERusenearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
CAUTION
DONOTusepediatric,lowflownasalcannulasoroxygen
maskswiththeoxygenconservingregulator.
DONOTblockthecannulaconnectionorkinkcannulatubing
whentheoxygenconservingregulatorisinuse.Thismay
damagetheoxygenconservingregulator.
1. Priortoeachuse,inspectproductforvisibledamage.DO
NOTuseifanydamageisfound.
ƽ WARNING
DONOTdirectflowofoxygenatanyperson,orflammable
materialwhenadjustingtheflowdial.
2. Turntheselectorswitchontheoxygenconserving
regulatortothe“0”OFFpositionbeforeopeningthe
cylindervalve.FIGURE4.2onpage21.
3. Toreducetheriskofrapidoxygenrecompressionandfire,
openthecylindervalveslowlyandcompletelysothe
pressuregaugemoves
slowlyasitindicatesthecylinder
pressure.