Sierra Monitor Corporation 4105 Stud Sensor User Manual


 
Model 4105 Gas Sensor Module (12/02)
Page: 4
Instruction Manual
4. OPERATION
4.1 Introduction
Under normal conditions the sensor module does not
require operator or technician intervention. The fol-
lowing are conditions under which the module requires
attention:
- Routine periodic calibration.
- Calibration after a high gas alarm.
- Sensor replacement on a planned schedule or
when a sensor failure occurs.
- Periodic cleaning as necessary.
- Unanticipated maintenance.
4.2 Signal Value
During normal operation the current loop of the sen-
sor module and the controller will be between 4 mA
indicating no presence of gas, and 20 mA indicating
that the full scale concentration of gas is present. The
signal value is proportional to the concentration of
gas present.