Performance Verification
4–8
THS710A, THS720A, THS730A & THS720P Service Manual
This test checks the bandwidth of channel 1.
1. Set up the TekScope instrument using the following steps:
SCO
PE
SA
VE/
RECALL
Recall Saved
Setup
Recall Factory
Setup
—
OK Recall
Factory
—
ACQUIRE Mode Average Set to 16
TRIGGER Coupling Noise Reject —
ME
AS
Gating & High-Low
Setup
High-Low Method Min/Max
Select Measrmnt Pk-Pk*
—
OK Select
Measrmnt
—
* You may need to press Select Page to display this selection.
2. Connect the TekScope instrument to the leveled sine wave generator as
shown below.
Leveled
sine wave
generator
Ch 1
Output
50 W feedthrough
terminator
BNC cable
Check Oscilloscope
Channel 1 Bandwidth