![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/b/2f/b2fd47cf-452f-4ec1-912c-73fb47177c4e/b2fd47cf-452f-4ec1-912c-73fb47177c4e-bg1c.png)
TDS 310, TDS 320 & TDS 350 Service Manual
1Ć11
Nominal Traits
Nominal traits are described using simple statements of fact such as Two,
identical" for the trait Input Channels, Number of," rather than in terms of
limits that are performance requirements.
TableĂ1Ć10:ăNominal Traits Ċ Signal Acquisition System
Name
Description
Bandwidth Selections 20 MHz and FULL
Digitizers, Number of Two, identical, digitized simultaneously
Digitized Bits, Number of 8 bits
1
Input Channels, Number of Two, identical, called CH 1 and CH 2
Input Coupling DC, AC, or GND
Ranges, Offset, All Channels Volts/Div Setting
2 mV/div - 99.5 mV/div
100 mV/div - 995 mV/div
1 V/div - 10 V/div
Offset Range
±1 V
±10 V
±100 V
Range, Position ±5 divisions
Range, Sensitivity
2
2 mV/div to 10 V/div
Rise Time TDS 310: 7 ns
TDS 320: 3.5 ns
TDS 350: 1.75 ns
TekProbe Interface Level one probe coding
1
Displayed vertically with 25 digitization levels (DLs) per division and 10.24 divisions dynamic range with zoom off. A DL is the
smallest voltage level change that the 8Ćbit AĆD Converter can resolve, with the input scaled to the volts/division setting of the channel
used. Expressed as a voltage, a DL is equal to 1/25 of a division times the volts/division setting.
2
The sensitivity ranges from 2 mV/div to 10 V/div in a 1-2-5 sequence of coarse settings. Between consecutive coarse settings, the
sensitivity can be finely adjusted with a resolution of 1% of the more sensitive setting. For example, between 50 mV/div and