Emerson Process Management 53eA Stud Sensor User Manual


 
MODEL 54eA SECTION 5.0
SOFTWARE CONFIGURATION
ITEM CHOICES FACTORY SETTINGS
CONFIGURE (continued)
C. Alarms (Section 5.7) (continued)
3. Alarm 1 setup (TPC)
a. Setpoint
if oxygen (ppm) -20 to 20 ppm 1 ppm
if oxygen (ppb) -999 to 999 ppb 100 ppb
if oxygen (% saturation) 0 to 200% 100%
if chlorine, monochloramine, or ozone -9999 to 9999 ppm 1.0 ppm
if pH -2.0 to 20.00 7.00
if temperature -5 to 130°C 25°C
b. Proportional 0 to 299.9% 100 %
c. Integral 0 to 2999 sec 0 sec
d. Derivative 0 to 299.9% 0.0%
e. Time period 10 to 2999 sec 30 sec
f. LRV (100% on)
if oxygen (ppm) -20 to 20 ppm 0 ppm
if oxygen (ppb) -999 to 999 ppb 0 ppb
if oxygen (% saturation) 0 to 200% 0%
if chlorine, monochloramine, or ozone -9999 to 9999 ppm 0 ppm
if pH -2.0 to 20.00 0.00
if temperature -5 to 130°C 0°C
g. URV (100% off)
if oxygen (ppm) -20 to 20 ppm 2 ppm
if oxygen (ppb) -999 to 999 ppb 100 ppb
if oxygen (% saturation) 0 to 200% 200%
if chlorine, monochloramine, or ozone -9999 to 9999 ppm 2 ppm
if pH -2.0 to 20.00 2.00
if temperature -5 to 130°C 100°C
h. Relay fault None, open, or closed None
4. Alarm 2 Control
a. Activation method Oxygen, chlorine, monochloramine, ozone, temperature, pH Oxygen
b. Control mode Normal or TPC Normal
5. Alarm 2 setup (normal)
a. Configuration Low, high, or off Low
Rest of alarm 2 setup is the same as alarm 1
6. Alarm 3 control and setup is the same as alarm 1
7. Alarm 4 control
Alarm Fault or off Fault
8. Feed limit timer
a. Feed limit Disable, alarm 1, alarm 2, or alarm 3 Disable
b. Timeout value 0 to 10,800 sec 600 sec
9. Interval timer
a. Select alarm Disable, alarm 1, alarm 2, or alarm 3 Disable
b. Interval time 0 to 999.9 hr 24.0 hr
c. Repeats 1 to 60 1
d. On time 0 to 2999 sec 120 sec
e. Off time 0 to 2999 sec 1 sec
f. Recovery time 0 to 999 sec 600 sec
TABLE 5-1. Program Settings List (continued)
Continued on the following page
19