Fluke PM-3390B Sander User Manual


 
4 - 44 COMMAND REFERENCE
Example 1:
Send
CONFigure:VOLTage:AC 0.6,(@2) Configures AC-RMS channel 2,
expected voltage 600 mV.
Send
INPut2:COUPling AC Channel 2 AC coupled.
Send
READ:AC? (@2) Initiates + fetches AC-RMS
value.
Read
<first measured AC-RMS value>
Send
READ:AC? (@2) Initiates + fetches AC-RMS
value.
Read
<second measured AC-RMS value>
Example 2:
Send
CONFigure:VOLTage:RISE:TIME (0.5),20,80,1E-2,(@2)
’Configures the rise time, expected voltage 0.5V,
’LOW ref. = 20%,
’HIGH ref. = 80%, expected time 0.01 seconds, channel 2.
Send
INPut2:COUPling DC Channel 2 becomes DC
coupled.
Send
READ:RISE:TIME? (@2) Initiates + fetches the rise time
of the signal on channel 2.
Read
<the measured rise time>
Send
FETCh:FALL:TIME? (@2) Fetches the fall time of the
signal on channel 2.
Read
<the measured fall time>