SENSe:SWEep:REALtime[:STATe]
Syntax:
SENSe:SWEep:REALtime[:STATe] <Boolean>
Query form:
SENSe:SWEep:REALtime[:STATe]?
Response:
0 | 1
0 Real-time mode switched off.
1 Real-time mode switched on.
Description:
Switches the ’real- time’ mode of the acquisition on or off. If the ’real-time’
sampling mode is switched on, the MTB range is limited to sequential sampling
from 250 nanoseconds through 200 seconds per division (for MTB ranges, refer
to the SENSe:SWEep:TIME command). In the analog mode error -221 "Settings
conflict;Digital mode required" is generated.
After a
*
RST command, the ’real-time’ mode is switched off.
Example:
Send
→ *RST Resets the instrument.
Send
→ SENSe:SWEep:REALtime ON Sets real-time sampling on.
Send
→ TRIGger:SOURce INTernal1;LEVel .1;SLOPe EITHer
Sets the following trigger settings: source = internal channel 1
level = 0.1V
slope = either pos. or neg.
Send
→ INITiate Initiates a single acquisition.
Send
→ READ:AC? Reads AC-RMS.
Read
← <AC-RMS voltage>
Front panel compliance:
The SENSe:SWEep:REALtime command is the remote equivalent of the front
panel REALTIME ONLY option of the TB MODE menu.