4 - 44 COMMAND REFERENCE
Example 1:
Send
→ CONFigure:VOLTage:AC 0.6,(@2) Configures AC-RMS channel 2,
expected voltage 600 mV.
Send
→ INPut2:COUPling AC Channel 2 AC coupled.
Send
→ READ:AC? (@2) Initiates + fetches AC-RMS
value.
Read
← <first measured AC-RMS value>
Send
→ READ:AC? (@2) Initiates + fetches AC-RMS
value.
Read
← <second measured AC-RMS value>
Example 2:
Send
→ CONFigure:VOLTage:RISE:TIME (0.5),20,80,1E-2,(@2)
’
’Configures the rise time, expected voltage 0.5V,
’LOW ref. = 20%,
’HIGH ref. = 80%, expected time 0.01 seconds, channel 2.
’
Send
→ INPut2:COUPling DC Channel 2 becomes DC
coupled.
Send
→ READ:RISE:TIME? (@2) Initiates + fetches the rise time
of the signal on channel 2.
Read
← <the measured rise time>
Send
→ FETCh:FALL:TIME? (@2) Fetches the fall time of the
signal on channel 2.
Read
← <the measured fall time>